This site is intended to be viewed with a Macromedia Flash Player.

Download Player 






Mycobacteriology | Parasitology | Quality Control | Clinical Lab Supplies | Mycology
 

 

 

 







 

Multimedia

Relative Centrifugal Force Calculator

The relative centrifugal force (RCF) used in laboratory sedimentation procedures is absolutely critical to organism recovery and diagnostic accuracy. Unfortunately the majority of laboratories mistakenly assume that RCF and RPM are equivalent. This failure can have serious consequences, especially in AFB specimen processing, where due to their buoyancy, mycobacteria are very difficult to sediment.

To help ensure our customers are processing their samples in the best possible way we have provided an easy to use calculator that allows for the conversion of RPM to RCF. To use this calculator simply measure the radius of your centrifuge in cm. This measurement should be from the center of the rotor to the extreme edge of the tube when the centrifuge is in motion. Once this measurement is taken, enter that number and the desired RCF (In the case of mycobacteria processing 3000xg is the proper RCF), and the calculator will return the appropriate RPM at which to set your centrifuge.



RCF = 0.00001118 * radius * RPM2
cm
g



. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . .

 

Back to News & Media

 


 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 
 